MRI Compatibility Testing for Accessories
Biodex Medical Systems’ MRI catalog items were tested at Stony Brook University Hospital for compatibility with their Phillips Achieva 3T MRI System. Terry Button, Ph.D., and Clyde Schlein, Director of Regulatory Affairs & Compliance at Biodex Medical Systems, were present.
Items were checked to see if they could be used near the bore of the MR Magnet. If there was no magnetic pull, they would then be determined to be acceptable for use with a 3T System.
MRI Compatibility Testing for Accessories
Biodex Medical Systems’ MRI catalog items were tested at Stony Brook University Hospital for compatibility with their Phillips Achieva 3T MRI System. Terry Button, Ph.D., and Clyde Schlein, Director of Regulatory Affairs & Compliance at Biodex Medical Systems, were present.
Items were checked to see if they could be used near the bore of the MR Magnet. If there was no magnetic pull, they would then be determined to be acceptable for use with a 3T System.
Test date: December 17, 2005
Following is the list of items and the results:
Model # | Description | Results |
240-100 | MR Stretcher | MR Conditional for 3T |
240-110 | MRI Stretcher with Fowler Positioning |
MR Conditional for 3T |
240-189 | I.V. Pole | MR Conditional for 3T |
240-075 | Safety Step Stool | MR Conditional for 3T |
240-091 | MRI I.V. Stand | MR Conditional for 3T |
240-805 | MRI Wheelchair Chair with I.V. Pole |
MR Conditional for 3T |
Chart revised: August 24, 2011
Clyde2005MRI compat testing for access.doc